Introducing RAININ Pipet-Lite™
Change Your Old Pipetting Habits
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Feature by feature, Pipet-Lite excels Volume setting Lock Snag-proof design for rapid volume adjustment. Helps you find and hold the zero position before aspiration. Ensures consistent sample pick up for improved user accuracy and precision. Same or better performance than Pipetman. “Hand-friendly“. Lets you use a lighter grip during pipetting and rest your hand between cycles, reducing grip force, fatigue, pipetting errors. Lighter springs and low-drag seals. Reduces thumb force. Reduces thumb impact and stress when ejecting tips. Fast, painless removal of ejector arm. No sore fingers. Pipet-Lite gives you the flexibility of choice to suit your pipetting needs. Select either traditional universal tip/shaft configuration (SL-) for use with universal tips. or LTS™ LiteTouch cylindrical system (L-) for maximum ergonomic benefits. LTS dramatically reduces tip ejection forces, only 0.6 kg compared to 4-10 kg with other pipettes. No extra cost for LTS pipetting system. |
|||||||||||||||||||||||||||||||
| RAININ Pipet-Lite™ Specifications | Gilson PIPETMAN® Specifications | ||||||
| Model | Volume µl |
Accuracy |
Precision Random Error (µl) |
Model | Volume µl | Accuracy Systematic Error (µl) |
Precision Random Error (µl) |
| SL-2 L-2 |
0.2 | ±0.024 | <0.012 | P2 F144801 |
0.2 | ±0.024 | <0.012 |
| 2 | ±0.03 | <0.014 | 2 | ±0.03 | <0.014 | ||
| SL-10 L-10 |
1 | ±0.025 | <0.012 | P10 F144802 |
1 | ±0.025 | <0.012 |
| 10 | ±0.1 | <0.04 | 10 | ±0.1 | <0.04 | ||
| SL-20 L-20 | 2 | ±0.15 | <0.04 | P20 F123600 |
2 | ±0.1 | <0.03 |
| 20 | ±0.2 | <0.06 | 20 | ±0.2 | <0.06 | ||
| SL-100 L-100 | 10 | ±0.35 | <0.1 | P100 F123615 |
20 | ±0.35 | <0.1 |
| 100 | ±0.8 | <0.15 | 100 | ±0.8 | <0.15 | ||
| SL-200 L-200 | 20 | ±0.5 | <0.2 | P200 F123601 |
50 | ±0.5 | <0.2 |
| 200 | ±1.6 | <0.3 | 200 | ±1.6 | <0.3 | ||
| SL-1000 L-1000 | 100 | ±3 | <0.6 | P1000 F123602 |
200 | ±3 | <0.6 |
| 1000 | ±8 | <1.5 | 1000 | ±8 | <1.5 | ||
| SL-5000 L-5000 | 500 | ±12 | <3 |
P5000 |
1000 | ±12 | <3 |
| 5000 | ±30 | <8 | 5000 | ±30 | <8 | ||
| SL-10ML L-10ML | 1ml | ±50 | <6 | P10ml F131201 |
1ml | ±30 | <6 |
| 10ml | ±60 | <16 | 10ml | ±60 | <16 | ||
N.B. In both cases these specifications are more stringent that those required by ISO 8655
ALL Pipet-Lite models are specified to within 10% of their nominal volume, ensuring you get great performance across the full volume range of the pipette.