Features include an electronic display, cycle counters, GLP security functions and autoclavable tip holder and ejector. Designed with robustness and precision in mind, it features exclusive enhancements and innovations that address the users concern for Repetitive Strain Injuries.
Features:
·Ultra Comfort: Ergonomic design makes it easier to grip, fits naturally in your hand, and requires less effort to use.
·Stress-free Pipetting: With the new design of the Pipetman Ultra, you produce less force while pipetting, which results in stress-free pipetting. Smoother pipetting every day!
·Natural Movement: Right- or left-handed? With the new three-position, natural-movement tip ejector, less force is required to eject tips ... resulting in less hand fatigue. No more sore thumbs!
·Quick Volume Setting: Set the volume quickly with only 10 turns from the lowest to the highest setting. Easy to set the volume—either with one hand using the thumb wheel (even when wearing gloves) or with two hands using the push button.
·Higher Tip Compatibility: Adjustable, patented tip ejector stroke for maximum compatibility with tips and for improved comfort when ejecting tips.
·Ultra Accuracy and Precision: Like all other Gilson pipettes, the Pipetman Ultra has been designed and manufactured to provide a robust, accurate, and precise instrument. Now, you can enjoy additional innovations that will be of practical assistance in your daily work.
·Exclusive Patented, Multi-functional Electronic Display: For greater readability with zero risk of parallax errors. Four-digit display is more accurate. Display shows status: volume, locked, or unlocked.
·Ultra Easy Maintenance: Prevention is the best cure. With proper care, your Pipetman Ultra will perform at the highest level of precision and accuracy, year after year. It is easy to dismantle and clean right in your own laboratory.
·Calibrate Easily with Only One Tool: Now, calibrating your Pipetman Ultra takes only a few minutes. The tool and its electronic display are a perfect pair for easy and secure calibration in your lab.
·Autoclavable: The parts of the Pipetman Ultra that normally come into contact with liquids (tip holder, tip ejector) are autoclavable.
·On-site Maintenance for Better Performance: The removable tip ejector allows easy access to the user-serviceable parts. The tip ejector is made of PVDF to avoid corrosion. The tip holder is in two parts for direct access to the piston and O-ring for easy maintenance easy maintenance and cleaning. Decontamination? Special applications? Breakage? Simply unscrew the lower part.
·Ultra GLP Features: Innovating to give you more security and traceability. Assist you in setting up GLPs down to pipetting level, giving you essential information.
Pipetting Workflow Control: Cycle Counters:
From the last time the volume was set—to count the number of cycles in the current run
Since the pipette was first used—to count the number of cycles for servicing purposes
Pipette Coding:
Personalize your pipette by writing your name or application on one of the color-coded labels (protected by the plastic cover). Or, use the bar code label from the warranty card.
Color-coded thumb wheeland push button identify model, with useful volume range shown on the push button.
GLP Security Functions:
Out of specification warning: the display flashes when selecting an out-of-specification volume
Lockable volume: with on-screen display for no risk of volume change during use
Specifications:
PIPETMAN U10 Specifications
Gilson Permissible Maximum Errors
ISO 8655 Maximum Permissible Errors
Model
Diamond Tips*
Part No.
Volume (µL)
Systematic Error (µL)
Random Error (µL)
Systematic Error (µL)
Random Error (µL)
U10
D10
DL10
F21022
1
5
10
+/- 0.025
+/- 0.075
+/-0.100
< 0.012
< 0.030
< 0.040
+/- 0.12
+/- 0.12
+/- 0.12
< 0.08
< 0.08
< 0.08
*Recommended tip for establishing maximum permissible errors (ISO8655 requirement).
A large range of suitable Gilson Diamond tips is available.