Gilson PIPETMAN P2 is a variable volume air displacement pipette for dispensing nanolitre volumes between 0.2µl and 2µl with precision accuracy and ease
Gilson PIPETMAN has been the worlds leading pipette for over 35 years providing the gold standard in accuracy, precision and robustness for scientists in all applications. Includes Dual Position Tip Ejector allowing use of Gilson Diamond D10 (short) and DL10 (long) tips and providing protection from cross contamination when using microtubes.
PIPETMAN P is a fully adjustable, air-displacement pipette with the selected volume shown on a digital indicator (volumeter).
Built to last: Constructed of stainless-steel and PVDF.
Unique "dry seal" technology avoids the need for grease to obtain an air-tight seal and guarantees long-term performance.
Thick handle in PVDF minimises hand warming effect during pipetting.
Rigorous quality control. Individually calibrated ... a Performance Check Report is included with each Pipetman, along with an engraved Identification Number for traceability
Easy maintenance, fully autoclavable tip-holder and tip-ejector.
P2 and P10 models are ideal for molecular biology techniques (PCR, DNA sequencing, etc.). Accurate and precise: P10 down to 1 µL and P2 down to 0.2 µL (0.1 µL with a good pipetting technique). PCR process is covered by U.S. patents owned by Hoffmann-La Roche Inc.
PIPETMAN P2 Specifications
Gilson Permissable Maximum Errors
ISO 8655 Maximum Permissable Errors
Model
Diamond Tips*
Part No.
Volume (µL)
Systematic Error (µL)
Random Error (µL)
Systematic Error (µL)
Random Error (µL)
P2
D10
F144801
0.2
+/- 0.024
< 0.012
+/- 0.08
< 0.04
DL10
0.5
+/- 0.025
< 0.012
+/- 0.08
< 0.04
2
+/-0.030
< 0.014
+/- 0.08
< 0.04
*Recommended tip for establishing maximum permissible errors (ISO8655 requirement).
A large range of suitable Gilson Diamond tips is available.