Gilson PIPETMAN Neo P10N is a variable volume air displacement pipette with reduced pipetting forces for dispensing nanolitre volumes between 1µl and 10µl with precision accuracy and ease
Pipette PIPETMAN Neo P10N 1-10uL
Gilson Pipetman Neo is an evolution of the classic Pipetman P that addresses the growing susceptibility of pipette users to repetitive strain injuries.
Characteristic for Pipetman P features such as: accuracy, precision and robustness remained unchanged. However, newly engineered springs provide a significant reduction of pipetting forces. P10N includes Dual Position Tip Ejector allowing use of Gilson Diamond D10 (short) or DL10 (long) tips and providing protection from cross contamination when using microtubes. PIPETMAN NEO is a fully adjustable, air-displacement pipette with the selected volume shown on a digital indicator (volumeter).
Built to last: Constructed of stainless-steel and PVDF
Unique “dry seal” technology avoids the need for grease to obtain an air-tight seal and guaranteed long-term performance
Thick handle in PVDF minimises hand warming effect during pipetting.
New colour-coded push-buttons allow for even easier identification of the different models and corresponding quality Diamond tips.
Significant reduction of pipetting and purge forces
Wide useful volume range: 10-100% of the nominal volume.
Rigorous quality control. A Performance Check Report is included with each Pipetman, along with an engraved Identification Number for traceability.
Easy maintenance, fully autoclavable tip-holder and tip-ejector
P2N and P10N models are ideal for molecular biology techniques (PCR, DNA sequencing, etc.) PCR process is covered by U.S. patents owned by Hoffmann-La Roche Inc.
Accurate and precise: P10N down to 1 µL and P2N down to 0.2µL.
PIPETMAN P10N Specifications
Gilson Permissable Maximum Errors
ISO 8655 Maximum Permissable Errors
Model
Diamond Tips*
Part No.
Volume (µL)
Systematic Error (µL)
Random Error (µL)
Systematic Error (µL)
Random Error (µL)
P10
D10
F144802
1
+/- 0.025
< 0.012
+/- 0.12
< 0.08
DL10
5
+/- 0.075
< 0.030
+/- 0.12
< 0.08
10
+/-0.100
< 0.040
+/- 0.12
< 0.08
* Recommended tip for establishing maximum permissible errors (ISO8655 requirement).
A large range of suitable Gilson Diamond tips available.